COMPARING METER AND LABORATORY RESULTS
Test results from the meter and laboratory are both
expressed in plasma-equivalent units. However, the
result you obtain from your meter may differ somewhat
from your laboratory result due to normal variation.
Meter results can be affected by factors and conditions
that do not affect laboratory results in the same way.
(See test strip package insert for typical accuracy
and precision data, and for important information on
limitations.) To make an accurate comparison between
meter and laboratory results, follow the guide-lines
below.
Before you go to the lab:
Perform a control solution test to make sure that the
meter is working properly.
It is best to fast for at least eight hours before doing
comparison tests.
Take your meter with you to the lab.
While at the lab:
Make sure that the samples for both tests (the meter test
and the lab test are taken and tested within 5 minutes
of each other).
Wash your hands before obtaining a blood sample.
Never use your meter with blood that has been
collected in a gray-top test tube.
Use fresh capillary blood only.